- Turn Chameleon to Standby.
- Switch off chiller and disconnect power line.
- Disconnect the coolant lines, ensure that they are labelled so that they are re-connected to the same positions.
- Dispose of the spent coolant from the chiller tank into the appropriate waste disposal bottle (stored next to the new coolant). IMPORTANT: it contains Ethylen Glycol, therefore it is harmful for the environment.
- Re-fill chiller with coolant ~325 ml.
- Re-connect the coolant lines to chiller.
- Re-connect power line and switch on chiller. Check that the flow is ok.
Coolant last changed on: Check lab calendar. Changed every 3 months by Erika / Istvan.
Perform a “data run” to check the laser performance every time you do a coolant exchange. As parameteres use every 10nm, 10sec. If you get close to 2W internal power it is fine.
- Turn ON the laser
- GLATransitionGUI opens the GUI for the laser
- Wait until the laser is stable
- The Tune window is the one for laser maintenance, from here you can run the “data run” (intuitive)
- Run a “sweap” before (this way the data run did not result in an error when trying to stop it)
- To start the “data run” choose the saving location (C:\coherent) and a file name (standard is : YYYY_MM_DD_datarun_User)
- A pop up window will appear, just write the engineer name (you can write the other infos if handy or if anything has changed from the previous time) and continue
- When the “data run” has gone through all wavelengths click “stop” (Roni says it used to stop by itself but it doesn't anymore)