I wanted to remove the detector head in order to examine how the objective mount is fixed in place
Turn off the Sutter MP285, PMTs and wide-field/2P servo motor control
Remove the objective!
Undo the connector for the electrical connection to the servo motor that switches the Dichroic from 2P to WF mode.
Remove 4 hex bolts holding the whole detector head and objective assembly to the Z motor (extreme right of our current setup)
To remove the objective mount itself
remove the detector (PMT) dichroic assembly by unscrewing the 2 hex bolts on the bottom plate using the smaller allen key
You can now access two similar bolts that screw through to the stainless steel objective mount
To release the objective mount, remove 2 philips head screws that attach it to a black 1 x 3 cm supporting strut
note that these screws are different lengths
when I removed them the longer one was in the middle of the fitting, but I think the reverse makes more sense
Careful you don't destroy any screws (especially the philips head screws) by using the wrong tool or forcing anything
Keep the screws safe
It can be a good idea to screw them back into their holes if you have something disassembled for a while