Table of Contents

Using the 2-photon microscope

Closing the main shutter of the laser

There are two ways to check and close the main shutter of the laser.

  1. preferred method Use the Chameleon Vision interface on the Windows computer (GLAT transit software icon link on Desktop if not open). In the system control tab of Chameleon Vision, check that the shutter is closed. If the shutter is open (and the laser is on or on stand-by), you see a yellow triangle. If that is the case, close the shutter. - since moving to 3N390 the software interface can't detect the laser, only the spectrophotometer, main shutter is controlled by button on laser controller
  2. On the laser controller, there is a button 'Shutter open' and a LED that turns on when the shutter is open. Press the button to close the main shutter.

Imaging with the 2-photon microscope

  1. Turn on the Windows computer on (login: Jefferis Lab with labpassword).
  2. Open Chameleon Vision (GLAT transit icon on Desktop). - currently only reads spectrophotometer data, not essential
  3. Switch the laser from 'stand-by' to 'on'. (The main shutter should be still closed)
  4. Turn on the imaging hardware on the rack with the PC screen:
  5. Turn on the MP-285 micromanipulator controller (and press the green button on the right - this used be necessary, not anymore).
  6. Turn on the MDR-6 galvo scanner controller
  7. Check that the two channels on the PS-2 are set to 'off'.
  8. Turn on the PS-2 PMT amplifier
  9. Turn on the Pockels cell (under the table, Conoptics box)
  10. Turn on the epi light pathway:
    1. Turn on the LED controller using the Thorlabs cube knob on the table (it says Blue light).
    2. Open the shutter on the Olympus filter wheel.
    3. Switch both servomotor controller to wide-field (WF). (next to keyboard of the PC)
  11. Put your sample under the objective with minimum amount of solution needed to avoid spills on the equipment.
  12. Check that the sample is lower than the objective.
  13. Turn off the main light in the room.
  14. Turn on the sign 'Laser in use'.
  15. Locate your sample:
    1. Lower the objective to about 2mm above the sample by eye.
    2. Ensure the objective is as well centered on the sample as possible.
    3. Add a drop of saline solution or water.
    4. Open MicroManager on the Mac computer, and load the config file of the Grasshopper camera (should be default)
    5. Click on Live.
    6. Use the XYZ controller to move the objective and locate your sample. The controller has a fast and slow mode. Press the red button '' to switch to slow mode; an LED will light up in slow mode.
    7. Focus in the middle of your sample.
  16. Turn off epi light pathway:
    1. Turn off LED controller using knob.
    2. Close shutter on filter wheel.
    3. Remove any objects that would block the hood from closing and CLOSE THE HOOD.
    4. Switch both servomotor controller to '2P'.
  17. On the Windows computer, open Matlab, type and run scanimage.
  18. Select MachineDataFile_Pockels.m which will run a Pockels Cell calibration and open a GUI.
  19. Open the laser light pathway:
    1. Open the main shutter on the laser controller / in Chameleon GUI
    2. Switch the two channels of the PS-2 to on. Indicated values should be between 0.400 and 0.450.
  20. To view your sample, click on 'FOCUS'.
  21. check the power meter standing on the laser, this indicates the laser intensity permitted to go through by the Pockels (that reaches your sample)
  22. Move the objective 80 microns up to locate and focus on your sample. The focus of the laser light is slightly higher than the focus of the epi light. However, be VERY careful when lowering the objective to avoid touching the sample or anything with the objective, always operate in slow mode while using the 2P. Be aware of the anatomy of your sample, e.g. if you reach the esophagus of the fly you shouldn't go too much further down.
  23. When you are happy with the imaging plane, PMT gain, and laser power, click Abort to terminate the Focus acquisition on ScanImage.
  24. In ScanImage select a folder to save files to, otherwise it will abort any effort to do this
  25. To save images, click on 'GRAB'. Enter a name for the images (e.g. YYYYMMDD_something_).
  26. For external triggering tick the box External Trigger and send trigger pulse from Igor (Mac) to start Grab or Loop acquisition
  27. Igor can send a trigger by running e.g. Preview (and selecting an ODD file, depending on your settings)
  28. you can select 'LOOP' instead of 'GRAB' to image multiple sweeps; these can be all triggered externally by running a multi-sweep protocol in Igor

Switching the 2-photon microscope off after imaging

  1. Close the main shutter of the laser.
  2. Switch off both channels on the PS-2.
  3. Switch light path from 2P to wide field with servo-motor controller
  4. Lift the hood until the click.
  5. Turn off 'Laser in use' sign.
  6. Turn on main light in room.
  7. Move objective away from sample using the x-y-z controller.
  8. Remove objective with lever - currently not used at 2P1 to have more space at the side of the camera
  9. Remove saline from sample to avoid spills when removing the sample.
  10. Remove sample.
  11. Clean objective :
    1. Rinse it abundantly with water from all sides while holding a beaker underneath the objective to collect the water.
    2. Wipe the objective with a sheet of Thorlabs objective paper. No scraping, only a single wipe in one direction.
    3. Repeat the rinse and wipe with a clean objective paper.
  12. Close matlab.
  13. Switch off PS-2.
  14. Switch off MDR-6.
  15. Switch off MP-285.
  16. Switch off pockel-cell.
  17. Switch laser to 'stand-by'.
  18. Copy data from computer to hard-disk and delete it from computer.

Configuration

Chameleon Viewer from GLAT transit

Scan Image ++

Maintenance

Trouble shooting

Objective does not move
  1. Check whether the x-y-z controller is in slow or fast mode.
  2. Check that the MP-285 indicates x-y-z positions. If it does not, press the green button again. If that does not help, restart ScanImage++ or matlab and press the green button on the MP-285.
No image in epi light
  1. Check that the epi pathway is fully open:
    1. LED controller switched on.
    2. Shutter on filter wheel is open.
    3. Both servomotor controllers are switched to 'WF'.
  2. Check that your sample is still there.
  3. Change the focal plane by moving objective. Be careful when going down.
No image in 2-photon light
  1. Check that your laser light pathway is open:
    1. Main shutter is open.
    2. imaging shutter is open (clicking sound when you press FOCUS)
    3. Both channels on the PS-2 are switched on.
    4. Both mirrors are set to 2P on the servo-motor controller
    5. Laser is on (not in 'stand-by').
  2. Check that your objective actually moves.
  3. Check that you have sufficient laser power (power meter is placed after Pockels)
  4. The laser might be misaligned, alignment requires training
  5. Change the focal plane. Move downwards and upwards but be careful with the objective!
  6. Switch back to epi light and check that your sample is still in place:
    1. Close laser main shutter.
    2. Turn both channels on PS-2 off.
    3. Open hood and return to epi light pathway.
    4. Check the sample is in place and in focus.

Components of 2-photon microscope