Turn on the Windows computer on (login: Jefferis Lab with labpassword).
Open Chameleon Vision (GLAT transit icon on Desktop). - currently only reads spectrophotometer data, not essential
Switch the laser from 'stand-by' to 'on'. (The main shutter should be still closed)
Turn on the imaging hardware on the rack with the PC screen:
Turn on the MP-285 micromanipulator controller (and press the green button on the right - this used be necessary, not anymore).
Turn on the MDR-6 galvo scanner controller
Check that the two channels on the PS-2 are set to 'off'.
Turn on the PS-2 PMT amplifier
Turn on the Pockels cell (under the table, Conoptics box)
Turn on the epi light pathway:
Turn on the LED controller using the Thorlabs cube knob on the table (it says Blue light).
Open the shutter on the Olympus filter wheel.
Switch both servomotor controller to wide-field (WF). (next to keyboard of the PC)
Put your sample under the objective with minimum amount of solution needed to avoid spills on the equipment.
Check that the sample is lower than the objective.
Turn off the main light in the room.
Turn on the sign 'Laser in use'.
Locate your sample:
Lower the objective to about 2mm above the sample by eye.
Ensure the objective is as well centered on the sample as possible.
Add a drop of saline solution or water.
Open MicroManager on the Mac computer, and load the config file of the Grasshopper camera (should be default)
Click on Live.
Use the XYZ controller to move the objective and locate your sample. The controller has a fast and slow mode. Press the red button '' to switch to slow mode; an LED will light up in slow mode.
Focus in the middle of your sample.
Turn off epi light pathway:
Turn off LED controller using knob.
Close shutter on filter wheel.
Remove any objects that would block the hood from closing and CLOSE THE HOOD.
Switch both servomotor controller to '2P'.
On the Windows computer, open Matlab, type and run scanimage.
Select MachineDataFile_Pockels.m which will run a Pockels Cell calibration and open a
GUI.
Open the laser light pathway:
Open the main shutter on the laser controller / in Chameleon
GUI
Switch the two channels of the PS-2 to on. Indicated values should be between 0.400 and 0.450.
To view your sample, click on 'FOCUS'.
check the power meter standing on the laser, this indicates the laser intensity permitted to go through by the Pockels (that reaches your sample)
Move the objective 80 microns up to locate and focus on your sample. The focus of the laser light is slightly higher than the focus of the epi light. However, be VERY careful when lowering the objective to avoid touching the sample or anything with the objective, always operate in slow mode while using the 2P. Be aware of the anatomy of your sample, e.g. if you reach the esophagus of the fly you shouldn't go too much further down.
When you are happy with the imaging plane, PMT gain, and laser power, click Abort to terminate the Focus acquisition on ScanImage.
In ScanImage select a folder to save files to, otherwise it will abort any effort to do this
To save images, click on 'GRAB'. Enter a name for the images (e.g. YYYYMMDD_something_).
For external triggering tick the box External Trigger and send trigger pulse from Igor (Mac) to start Grab or Loop acquisition
Igor can send a trigger by running e.g. Preview (and selecting an ODD file, depending on your settings)
you can select 'LOOP' instead of 'GRAB' to image multiple sweeps; these can be all triggered externally by running a multi-sweep protocol in Igor