added Multiclamp 700B from Jonny's / Erika's rig; 700A with headstages put there
Marco lent us their Picospritzer, to be triggered from ITC-18 TTL2
added a Luigs Neumann manipulator to target pipettes (the other one has been in use for delivering mechanic / taste stimuli to flies, also it has a very suspicious Y axis, handle with care)
installed Multiclamp 700B software + Janelia's wavesurfer on PC to control electrode (only one channel)
Primary and Secondary signals from Ch1 electrode go to Dev1 (NI PCI6323) AI6 and AI7 respectively; Command to electrode comes from same NI board's AO1
Piezo disabled (would need a second breakout box, or even better a second NI card - see below) to have enough AO channels
Scanimage Machine Data Files for Piezo changed to have AO3 as Piezo control channel to avoid initialising Scanimage with AO1 as Piezo control and kill the headstage - to access AO3 we need another breakout box
Recorded baseline noise with model cell in wavesurfer; power cables to the two power meters for laser intensity caused 10mV noise each, these will run from battery
After this one gets a fairly nice baseline with the imaging box closed, but turning on the PMTs adds high frequency noise